Select function ”Pump down”. ”Speed” ”HI” is recommended. For gel dryers set a lower speed if
necessary (if the gels tend to break).
Set ”Minimum” to prevent volatile components to evaporate. The process is stopped automatically as
soon as ”Minimum” is reached.
Set a process time (”Duration”) if necessary.
Start process by pressing key ”START/STOP”.
alternatively:
Select function ”Vac control” to dry at a predetermined vacuum level.
Set ”Set vacuum” to the evaporation vacuum of the solvent.
Set a process time (”Duration”) if necessary.
Start process by pressing key ”START/STOP”.
Assembly of a vacuum system
Assemble vacuum connection lines between vacuum pump, CVC 3000 and vacuum application.
Assemble the vacuum connection line between the controller and the vacuum application, if the con-
troller is not part of a pumping unit.
Assemble electrical connections.
Connect coolant if necessary.
Vacuum for ltration and suction
Select function ”Pump down”.
If necessary set value for ”Speed” (high or low pumping speed). Reduce the speed in case of fast
sedimentation (formation of a lter cake).
Set ”Set vacuum” to a value which still does not lead to evaporation of the solvent.
Start process by pressing key ”START/STOP”.
alternatively:
Select function ”Vac control”.
Set ”Set vacuum” to a value which still does not lead to evaporation of the solvent.
If necessary set value for ”Speed” (high or low pumping speed). Reduce the speed in case of fast
sedimentation (formation of a lter cake).
Set ”Maximum” to switch off the pump at the end of the process or in case of a lter cake crack (pres-
sure increase).
Start process by pressing key ”START/STOP”.
Tip for ltration: Adjust preset pressure to a value well above the boiling pressure of the solvent
(e.g. water >>20 mbar). Set the maximum pressure to e.g. 500 mbar. Once the ltration has nished,
the pressure increases and the pump is switched off automatically.
Vacuum for gel dryer, drying chambers and vacuum concentrators